Figure 8.14a shows the impulse response of a classical guitar body sampled at kHz. It was determined empirically that at least the first msec of this impulse response needs to be stored in the excitation table to produce a high quality synthetic guitar. Figure 8.14b shows the same impulse response after factoring out a single resonating mode near Hz (the main Helmholtz air mode). A close-up of the initial response is shown in Fig. 8.15. As can be seen, the residual response is considerably shorter than the original.
Figure 8.16a shows the guitar-body amplitude response, and Fig. 8.16b shows the response after the main Helmholtz air mode is removed by inverse filtering with a two-pole, two-zero filter. Figure 8.18 shows the same thing but with only a two-zero inverse filter; in this case the overall spectral shape is more affected.
Figure 8.19a shows the measured guitar-body amplitude response after warping to an approximate Bark frequency axis. Figure 8.19b shows the Bark-warped amplitude response after the main Helmholtz air mode is removed by inverse filtering. Fig. 8.20 shows the low-frequency close-up. The warped amplitude response was computed as the FFT of the impulse response of the FIR filter given by the original impulse response with each unit delay being replaced by a first-order allpass filter, as originally suggested in [337] and described further in [230].
Fig. 8.21 shows the corresponding results if a two-zero inverse filter is used rather than a two-pole, two-zero inverse filter, i.e., without the ``isolation poles,'' ( in Eq.(8.18) below). In this case, there is an overall ``EQ'' boosting high frequencies and attenuating low frequencies. However, comparing Figs. 8.18b and 8.21b, we see that the global EQ effect is less pronounced in the Bark-warped case. On the Bark frequency scale, it is much easier numerically to eliminate the main air mode.
The modal bandwidth used in the inverse filtering was chosen to be Hz which corresponds to a of for the main air mode. If the Bark-warping is done using a first-order conformal map [460], its inverse preserves filter order [432, pp. 61-67]. Applying the inverse warping to the parametric resonator drives its pole radius from in the Bark-warped plane to in the unwarped plane.
Note that if the inverse filter consists only of two zeros determined by the spectral peak parameters, other portions of the spectrum will be modified by the inverse filtering, especially at the next higher resonance, and in the linear trend of the overall spectral shape. To obtain a more localized mode extraction (useful when the procedure is to be repeated), we define the inverse filter as
Because the main air mode is extremely narrow, the probability of overflow can be reduced in fixed-point implementations by artificially dampening it. Reducing the of the main Helmholtz air mode from to corresponds to a decay time of about sec. This is consistent with the original desire to retain the first msec of the body impulse response.
For completeness, the Bark-warped impulse-responses are also shown in Figs. 8.22. Figure 8.22a shows the complete Bark-warped impulse response obtained by taking the inverse FFT of Fig. 8.19a, and Fig. 8.22b shows the shortened Bark-warped impulse response defined as the inverse FFT of Fig. 8.19b. We see that given a Bark-warped frequency axis (which more accurately represents what we hear), the time duration of the high-frequency modes is extended while the low-frequency modes are contracted in time duration. Thus, the modal decay times show less of a spread versus frequency. This also accounts for the reduced apparent shortening by the inverse filtering in the Bark-warped case.